arXiv 1902.01393q-bio.QMphysics.optics
收藏

无相机高通量3D成像流式细胞术

Cameraless High-throughput 3D Imaging Flow Cytometry

Yuanyuan Han, Rui Tang, Yi Gu, Alex Ce Zhang, Wei Cai, Violet Castor, Sung Hwan Cho, William Alaynick, Yu-Hwa Lo

摘要:

对理解细胞表型的巨大异质性的日益增长的需求推动了成像流式细胞术(IFC)的发展,其将流式细胞术的特征与荧光和明场显微镜相结合。IFC将流式细胞术的通量和统计优势与基于真实或计算图像以及常规流式细胞术指标离散测量事件的能力相结合。现有IFC系统的局限性在于,不管检测方法如何,仅获得二维(2D)细胞图像。在没有断层扫描三维(3D)分辨率的情况下,投影问题仍然存在:将3D信息折叠到2D图像上,限制了对细胞表型至关重要的点计数或共定位的可靠性。在这里,我们提出了一个解决方案的投影问题:三维成像流式细胞术(3D-IFC),一个高通量的三维细胞成像仪的基础上光学切片显微镜。我们结合正交光片扫描照明与我们以前的时空变换检测产生3D细胞图像重建从一个无相机的单像素光电探测器读出。我们进一步证明了这种能力,通过共同捕获的3D荧光和无标记的侧散射图像的单细胞在流中的速度为0.2 m s-1,对应于约500个细胞每秒的吞吐量与60,000体素(随后调整大小为106体素)为每个细胞图像在小于1微米的分辨率在X,Y和Z尺寸。需要改进的高通量成像工具来在免疫学、肿瘤学、细胞和基因治疗以及药物发现领域中识别表型-基因型异质性。

英文摘要:

Increasing demand for understanding the vast heterogeneity of cellular phenotypes has driven the development of imaging flow cytometry (IFC), that combines features of flow cytometry with fluorescence and bright field microscopy. IFC combines the throughput and statistical advantage of flow cytometry with the ability to discretely measure events based on a real or computational image, as well as conventional flow cytometry metrics. A limitation of existing IFC systems is that, regardless of detection methodology, only two-dimensional (2D) cell images are obtained. Without tomographic three-dimensional (3D) resolution the projection problem remains: collapsing 3D information onto a 2D image, limiting the reliability of spot counting or co-localization crucial to cell phenotyping. Here we present a solution to the projection problem: three-dimensional imaging flow cytometry (3D-IFC), a high-throughput 3D cell imager based on optical sectioning microscopy. We combine orthogonal light-sheet scanning illumination with our previous spatiotemporal transformation detection to produce 3D cell image reconstruction from a cameraless single-pixel photodetector readout. We further demonstrate this capability by co-capturing 3D fluorescence and label-free side-scattering images of single cells in flow at a velocity of 0.2 m s-1, corresponding to a throughput of approximately 500 cells per second with 60,000 voxels (resized subsequently to 106 voxels) for each cell image at a resolution of less than 1 micron in X, Y, and Z dimensions. Improved high-throughput imaging tools are needed to phenotype-genotype recognized heterogeneity in the fields of immunology, oncology, cell- and gene- therapy, and drug discovery.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑