Partial renal coverage in EVAR causes unfavorable renal flow patterns in an infrarenal aneurysm model
摘要:
目的:为了在腹主动脉瘤腔内修复术期间实现最佳密封区,覆膜支架织物近端的预期定位应尽可能靠近肾动脉的最尾侧边缘。一些覆膜支架在不透射线标记和近端织物边缘之间存在小偏移。因此,可能发生意外的部分肾动脉覆盖。本研究探讨了部分覆盖对肾血流模式和肾壁切应力的影响。研究方法:腹主动脉瘤的体外模型被用于在生理静息条件下使用2D粒子图像测速仪可视化脉动流。一个模型作为对照,两个模型使用Endurant覆膜支架植入支架,一个没有,一个部分覆盖肾动脉,1.3 mm的支架织物延伸超过标记(16\%面积覆盖)。分析了肾动脉壁面切应力、相对停留时间和返流的大小和振荡情况。结果如下:在两种支架模型中,沿肾尾动脉壁的区域存在较低且振荡的壁剪切应力,而对照模型中不存在。与没有肾脏覆盖的模型中的2mm长度相比,在具有7mm长度的部分覆盖的模型中存在具有非常低的壁剪切应力(<0.1Pa)的区域。肾动脉中的平均肾返流面积百分比从对照组(0.9%)逐渐增加至无肾脏覆盖(6.4%)和有肾脏覆盖(18.8%)的支架模型。结论:在该血流模型中,腹主动脉瘤腔内修复术后部分肾脏覆盖导致壁剪切应力的低和显著振荡,可能促进动脉粥样硬化和随后的肾动脉狭窄。因此,在血管内实践中,了解织物边缘和不透射线标记之间的器械相关偏移非常重要。
英文摘要:
Objective: To achieve an optimal sealing zone during EVAR, the intended positioning of the proximal end of the endograft fabric should be as close as possible to the most caudal edge of the renal arteries. Some endografts exhibit a small offset between the radiopaque markers and the proximal fabric edge. Unintended partial renal artery coverage may thus occur. This study investigates the consequences of partial coverage on renal flow patterns and wall shear stress. Methods: In-vitro models of an abdominal aortic aneurysm were used to visualize pulsatile flow using 2D particle image velocimetry under physiologic resting conditions. One model served as control and two models were stented with an Endurant endograft, one without and one with partial renal artery coverage with 1.3 mm of stent fabric extending beyond the marker (16\% area coverage). The magnitude and oscillation of wall shear stress, relative residence time and backflow in the renal artery were analyzed. Results: In both stented models, a region along the caudal renal artery wall presented with low and oscillating wall shear stress, not present in the control model. A region with very low wall shear stress (<0.1 Pa) was present in the model with partial coverage over a length of 7 mm, compared to a length of 2 mm in the model without renal coverage. Average renal backflow area percentage in the renal artery incrementally increased from control (0.9%) to the stented model without (6.4%) and with renal coverage (18.8%). Conclusion: In this flow model partial renal coverage after EVAR causes low and marked oscillations in wall shear stress, potentially promoting atherosclerosis and subsequent renal artery stenosis. Awareness of the device-dependent offset between the fabric edge and the radiopaque markers is therefore important in endovascular practice.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。