ALD涂层Planacon MCP-PMT的阳极电流饱和

Anode current saturation of ALD-coated Planacon MCP-PMTs

E. V. Antamanova, I. G. Bearden, E. J. Garcia-Solis, A. V. Harton, V. A. Kaplin, T. L. Karavicheva, J. L. Klay, Yu. A. Melikyan, D. V. Serebryakov, M. Slupecki, W. H. Trzaska

摘要:

我们测量并比较了ALD涂层Planacon MCP-PMT(XP 85112/A1-Q-L)与非ALD对应物(XP 85012/A1-Q)的特性。虽然后者表现出出色的性能,但ALD涂层传感器具有令人惊讶的低电流饱和水平(比预期低两个数量级)和极高的增益恢复时间(比预期高7个数量级以上)。我们怀疑这些问题可能是由于ALD工艺的意外副作用引起的。为了得出明确的结论,需要测试更多的样品,最好是来自不同的生产运行。如果我们的观察得到证实,这将意味着ALD涂层的MCP-PMT的严重技术挫折。

英文摘要:

We have measured and compared the characteristics of ALD-coated Planacon MCP-PMTs (XP85112/A1-Q-L) with their non-ALD counterparts (XP85012/A1-Q). While the later show excellent performance, the ALD-coated sensors have surprisingly low current saturation levels (~two orders of magnitude lower than expected) and extremely high gain recovery time (more than 7 orders of magnitude higher than expected). We suspect that these problems might be caused by the unexpected side-effects of the ALD process. To make a definite conclusion, more samples need to be tested, preferably from different production runs. If our observation were confirmed, it would mean a serious technological setback for ALD-coated MCP-PMTs.