arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 1805.09121physics.med-ph
收藏

光电容积描记器估测血压趋势和夜间下降

Estimating blood pressure trends and the nocturnal dip from photoplethysmograph

Mustafa Radha, Koen de Groot, Nikita Rajani, Cybele CP Wong, Nadja Kobold, Valentina Vos, Pedro Fonseca, Nikolaos Mastellos, Petra A Wark, Nathalie Velthoven, R… 展开作者

Mustafa Radha, Koen de Groot, Nikita Rajani, Cybele CP Wong, Nadja Kobold, Valentina Vos, Pedro Fonseca, Nikolaos Mastellos, Petra A Wark, Nathalie Velthoven, Reinder Haakma, Ronald M Aarts

摘要:

目的:在自由生活的个体中,使用腕戴式光电体积描记(PPG)传感器和深度神经网络,评价根据24小时血压趋势估计夜间收缩压(SBP)下降的方法,将深度神经网络与传统机器学习和非机器学习基线进行比较。方法:106名健康个体佩戴腕戴式PPG传感器226天,在此期间,使用24小时动态血压监测仪获得5111个血压(BP)参考值,并与PPG传感器数据相匹配。从PPG波形中提取基于心率变异性和脉搏形态的特征。长期和短期记忆(LSTM)网络,密集网络,随机森林和线性回归模型进行了训练和评估,以跟踪BP趋势以及SBP下降估计的能力。主要结果:使用深度LSTM神经网络获得了估计SBP下降的最佳性能,均方根误差(RMSE)为3.12 $\pm$ 2.20 $Δ$ mmHg,相关性为0.69 $(p=3*10^{-5})$。该下降来自血压的趋势估计值,收缩压的RMSE为8.22 $\pm$ 1.49 mmHg,舒张压(DBP)的RMSE为6.55 $\pm$ 1.39 mmHg。虽然其他模型在跟踪相对BP方面具有类似的性能,但它们在SBP下降方面的性能不如LSTM。意义:这项工作提供了第一个证据,夜间SBP下降,一个高度预后的临床参数的非侵入性估计。它也是第一个在自由生活的个体中不受约束的24小时测量的大型数据集中评估非侵入性BP测量的方法,并为LSTM模型在该领域的实用性提供了证据。

英文摘要:

Objective: Evaluate a method for the estimation of the nocturnal systolic blood pressure (SBP) dip from 24-hour blood pressure trends using a wrist-worn photoplethysmography (PPG) sensor and a deep neural network in free-living individuals, comparing the deep neural network to traditional machine learning and non-machine learning baselines. Approach: A wrist-worn PPG sensor was worn by 106 healthy individuals for 226 days during which 5111 reference values for blood pressure (BP) were obtained with a 24-hour ambulatory BP monitor and matched with the PPG sensor data. Features based on heart rate variability and pulse morphology were extracted from the PPG waveforms. Long- and short term memory (LSTM) networks, dense networks, random forests and linear regression models were trained and evaluated in their capability of tracking trends in BP, as well as the estimation of the SBP dip. Main results: Best performance for estimating the SBP dip were obtained with a deep LSTM neural network with a root mean squared error (RMSE) of 3.12$\pm$2.20 $Δ$mmHg and a correlation of 0.69 $(p=3*10^{-5})$. This dip was derived from trend estimates of BP which had an RMSE of 8.22$\pm$1.49 mmHg for systolic and 6.55$\pm$1.39 mmHg for diastolic BP (DBP). While other models had similar performance for the tracking of relative BP, they did not perform as well as the LSTM for the SBP dip. Significance: The work provides first evidence for the unobtrusive estimation of the nocturnal SBP dip, a highly prognostic clinical parameter. It is also the first to evaluate unobtrusive BP measurement in a large data set of unconstrained 24-hour measurements in free-living individuals and provides evidence for the utility of LSTM models in this domain.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑