基于优化的光栅干涉相位步进序列评价及机械安装不稳定性分析

Optimization based evaluation of grating interferometric phase stepping series and analysis of mechanical setup instabilities

Jonas Dittmann, Andreas Balles, Simon Zabler

摘要:

可由X射线光栅干涉仪访问的衍射对比度模态不直接成像,而是必须从在干涉仪的光栅的变化的相对位置处获取的一系列图像中发生的正弦信号变化来推断。在这些相位步进系列的采集中涉及的绝对空间平移通常位于仅几百纳米的范围内,因此,如果不考虑的话,小至10纳米的定位误差将已经转化为最终图像中百分之一到百分之十的信号不确定性。传统上,相位步进系列中的相对光栅位置被认为是分析的输入参数,并且对于通常采用的快速傅立叶变换,需要在光栅周期的倍数上等距分布。相比之下,对采样位置具有宽松要求的更一般的基于优化的方法通常被认为太耗时。在下文中,提出了一种快速收敛优化方案,其同时确定相位步进曲线的参数以及步进光栅的实际执行的运动,还包括通常被忽略的错误旋转运动。虽然发现仅沿步进方向的平移误差的校正对于减少图像伪影是足够的,但是也检测围绕所有轴的微小旋转的可能性被证明是用于系统校准和监测的有价值的工具。所提供的算法的简单性,特别是当只考虑平移误差时,使得它也非常适合作为大型图像系列的标准评估过程。

英文摘要:

The diffraction contrast modalities accessible by X-ray grating interferometers are not imaged directly but have to be inferred from sine like signal variations occurring in a series of images acquired at varying relative positions of the interferometer's gratings. The absolute spatial translations involved in the acquisition of these phase stepping series usually lie in the range of only a few hundred nanometers, wherefore positioning errors as small as 10nm will already translate into signal uncertainties of one to ten percent in the final images if not accounted for. Classically, the relative grating positions in the phase stepping series are considered input parameters to the analysis and are, for the Fast Fourier Transform that is typically employed, required to be equidistantly distributed over multiples of the gratings' period. More general optimization based approaches with relaxed requirements on the sampling positions are often deemed too time consuming in contrast. In the following, a fast converging optimization scheme is presented simultaneously determining the phase stepping curves' parameters as well as the actually performed motions of the stepped grating, including also erroneous rotational motions which are commonly neglected. While the correction of solely the translational errors along the stepping direction is found to be sufficient with regard to the reduction of image artifacts, the possibility to also detect minute rotations about all axes proves to be a valuable tool for system calibration and monitoring. The simplicity of the provided algorithm, in particular when only considering translational errors, makes it well suitable as a standard evaluation procedure also for large image series.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑