通过 falsification 减少模型不变安全保持控制中的保守性:丙泊酚麻醉的模型不变安全性系统

Reducing Conservatism in Model-Invariant Safety-Preserving Control of Propofol Anesthesia Using Falsification

University of British Columbia(不列颠哥伦比亚大学)

Mahdi Yousefi, Klaske van Heusden, Ian M. Mitchell, J. Mark Ansermino, Guy A. Dumont

AI总结:

本文提出一种基于反馈数据进行模型 falsification 的模型不变安全系统,用于减少麻醉闭环控制中的保守性,通过保持丙泊酚血浆浓度和血压在安全范围内来确保患者安全。

摘要:

本文提供了一种形式化的模型不变安全系统,用于闭环麻醉,该系统利用测量数据反馈进行模型 falsification 以减少保守性。安全系统在患者对丙泊酚的反应存在不确定性时,仍能保持预测的丙泊酚血浆浓度和血压在安全范围内。模型不变形式验证用于规范安全系统。该技术需要多模型描述模型不确定性。模型不变验证考虑了不确定系统的所有可能动态,因此所得的安全系统可能对不表现出最坏情况动态响应的系统而言较为保守。本文采用模型 falsification 来减少模型不变安全系统的保守性。如果模型集中的成员预测与不确定系统的测量响应之间存在差异,则这些模型将被 falsified,从而减少模型不确定性。我们证明了在模型不变安全系统中包含 falsification 可以减少安全系统的保守性。

英文摘要:

This work provides a formalized model-invariant safety system for closed-loop anesthesia that uses feedback from measured data for model falsification to reduce conservatism. The safety system maintains predicted propofol plasma concentrations, as well as the patient's blood pressure, within safety bounds despite uncertainty in patient responses to propofol. Model-invariant formal verification is used to formalize the safety system. This technique requires a multi-model description of model-uncertainty. Model-invariant verification considers all possible dynamics of an uncertain system, and the resulting safety system may be conservative for systems that do not exhibit the worst-case dynamical response. In this work, we employ model falsification to reduce conservatism of the model-invariant safety system. Members of a model set that characterizes model- uncertainty are falsified if discrepancy between predictions of those models and measured responses of the uncertain system is established, thereby reducing model uncertainty. We show that including falsification in a model-invariant safety system reduces conservatism of the safety system.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑