磁性纳米粒子热疗吸收功率的非量热法测定
Non-calorimetric determination of absorbed power during magnetic nanoparticle based hyperthermia
摘要:
纳米磁热疗技术是近年来研究的热点,具有广阔的应用前景。一个主要的挑战是确定在体内条件下的耗散功率,传统的方法要么是侵入性的,要么是不准确的。我们提出了一种非量热的方法,产生在热疗过程中吸收的热量:它是基于精确测量的谐振射频电路,这是用于照射的品质因数变化。该方法提供了实时的吸收功率,而不需要监测作为时间的函数的样品温度。因此,它不存在通常量热法的非绝热加热条件所引起的问题。我们通过比较耗散功率与传统的量热测量来验证该方法。我们提出了两种类型的谐振器具有非常不同的填充因子的验证:螺线管和所谓的鸟笼线圈。后者是体积线圈,通常用于体内条件下的磁共振成像(MRI)。因此,所提出的方法允许有效地组合MRI和热疗,并且因此容易适应于现有的成像硬件。
英文摘要:
Nanomagnetic hyperthermia (NMH) is intensively studied with the prospect of cancer therapy. A major challenge is to determine the dissipated power during in vivo conditions and conventional methods are either invasive or inaccurate. We present a non-calorimetric method which yields the heat absorbed during hyperthermia: it is based on accurately measuring the quality factor change of a resonant radio frequency circuit which is employed for the irradiation. The approach provides the absorbed power in real-time, without the need to monitor the sample temperature as a function of time. As such, it is free from the problems caused by the non-adiabatic heating conditions of the usual calorimetry. We validate the method by comparing the dissipated power with a conventional calorimetric measurement. We present the validation for two types of resonators with very different filling factors: a solenoid and a so-called birdcage coil. The latter is a volume coil, which is generally used in magnetic resonance imaging (MRI) under in vivo condition. The presented method therefore allows to effectively combine MRI and thermotherapy and is thus readily adaptable to existing imaging hardware.