DARHT轴1补偿罐D点校准
DARHT Axis 1 Compensation Can D-dot Calibration
摘要:
我们详细介绍了校准的D点探头上使用的洛斯阿拉莫斯国家实验室DARHT轴1 LIA补偿罐(CC)。我们希望这将不仅作为这次校准的记录,而且还作为如何在部署D点的其他系统上执行类似校准的指南。虽然是一个简单的测量,但其困难在于,在校准期间必须使用现场使用的相同几何形状,最终目标是测量探针传感器和想要测量其电压的组件之间的电容。由于其与电压的线性关系,这可以在比励磁期间更低的电压下完成,唯一的警告是该较低电压脉冲必须仍然提供足够的信噪比。在对我们的结果进行简要总结之后,我们包括对补偿罐、D点探针及其操作的描述。然后,我们提供了该探头对补偿罐内被测高压套管的电容的3D仿真结果。最后,我们描述了我们的校准设置和讨论我们的结果。
英文摘要:
We detail the calibration of the D-dot probes used on the Los Alamos National Laboratory DARHT Axis 1 LIA compensation cans (CCs). We hope this will serve not only as a record of this calibration, but also as a guide on how to perform similar calibrations on other systems where D-dots are deployed. Although a simple measurement, its difficulty lies in the fact that the same geometry used when fielded must be used during calibration, the ultimate goal being to measure the capacitance between the probe sensor and the component whose voltage one intends to measure. Because of its linearity with voltage, this can be done at a lower voltage than during fielding, the only caveat being that this lower voltage pulse must still provide enough signal to noise. After a brief summary of our results, we include a description of a compensation can, the D-dot probe and its operation. We then provide 3D simulation results of the capacitance of this probe to the high voltage bushing under test inside of the compensation can. Finally, we describe our calibration setup and discuss our results.