使用聚焦X射线束的X射线发光计算机断层成像
X-ray luminescence computed tomography using a focused X-ray beam
摘要:
由于电子倍增电荷耦合器件(EMCCD)相机装置的X射线光子利用率低和测量灵敏度低,基于准直器的窄束X射线发光计算机断层成像(XLCT)通常需要较长的测量时间。本文首次报道了一种基于单束光纤和光电倍增管的聚焦X射线束XLCT成像系统。使用具有多毛细管透镜的X射线管来产生聚焦的X射线束,其X射线光子密度比准直的X射线束大1200倍。采用光纤束收集体模表面上的发射光子并将其传递到PMT。总测量时间缩短至12.5分钟。对于单纤维束和六纤维束情况的数值模拟,我们能够成功地重建六个目标。对于体模实验,两个目标的边缘到边缘的距离为0.4 mm和中心到中心的距离为0.8 mm的测量设置与一个单一的光纤束和PMT成功地重建。
英文摘要:
Due to the low X-ray photon utilization efficiency and low measurement sensitivity of the electron multiplying charge coupled device (EMCCD) camera setup, the collimator based narrow beam X-ray luminescence computed tomography (XLCT) usually requires a long measurement time. In this paper, we, for the first time, report a focused X-ray beam based XLCT imaging system with measurements by a single optical fiber bundle and a photomultiplier tube (PMT). An X-ray tube with a polycapillary lens was used to generate a focused X-ray beam whose X-ray photon density is 1200 times larger than a collimated X-ray beam. An optical fiber bundle was employed to collect and deliver the emitted photons on the phantom surface to the PMT. The total measurement time was reduced to 12.5 minutes. For numerical simulations of both single and six fiber bundle cases, we were able to reconstruct six targets successfully. For the phantom experiment, two targets with an edge-to-edge distance of 0.4 mm and a center-to-center distance of 0.8 mm were successfully reconstructed by the measurement setup with a single fiber bundle and a PMT.