氦-3正比计数器的空间分辨率和Gamma-ray分辨特性

Characterisation of the Spatial Resolution and the Gamma-ray Discrimination of Helium-3 Proportional Counters

Edoardo Rossi

摘要:

讨论了几种氦-3正比管的能量分辨率。氦-3探测器的能量分辨率与可以实现的伽马射线和中子之间的区分密切相关。我们还讨论了不同的贡献的能量分辨率,这些提供有关如何氦-3计数器操作的信息。我们测量了位置敏感氦-3探测器的空间分辨率,以量化和了解这种技术可以实现的极限。伽马射线灵敏度是将伽马射线脉冲误计为中子脉冲的效率。这个量对于确定探测器的性能至关重要,文献中描述的伽马射线灵敏度测量都没有涉及氦-3探测器。我们测量了几个氦-3探测器的伽马射线灵敏度,为替代技术提供了一个有用的比较术语。我们研究了不同的贡献,引起的伽玛射线探测器的灵敏度。为了更好地理解这些贡献,我们进行了几次蒙特卡罗模拟。为了提高γ射线探测灵敏度,我们测试了利用脉冲形状鉴别(PSD)技术抑制γ射线脉冲的可行性。在证明其可行性的同时,我们量化了氦-3管中伽马射线抑制的幅度。还研究了PSD技术在氦-3探测器上的实现,以便为PSD在硼-10探测器上的实现设置一个可能的起点。

英文摘要:

We discuss over the energy resolution we measured for several Helium-3 proportional tubes. The energy resolution of a Helium-3 detector is strictly related to the discrimination between gamma-rays and neutrons that can be achieved. We also discuss the different contributions to the energy resolution; these provide information about how the helium-3 counters operate. We measured the spatial resolution of position sensitive Helium-3 detectors in order to quantify and understand the limit that can be achieved with this technology. The gamma-ray sensitivity is the efficiency for miscounting a gamma-ray pulse as a neutron pulse. This quantity is crucial to determine the performance of the detectors.None of the gamma-ray sensitivity measurements described in the literature involves Helium-3 detectors. We measured the gamma-ray sensitivity for several Helium-3 detectors, providing a useful term of comparison for the alternative technologies. We investigated the different contributions that give rise to the gamma-ray sensitivity in a detector. In order to better understand these contributions we performed several Monte Carlo simulations. In order to improve the gamma-ray sensitivity, we tested the feasibility of gamma-ray pulse rejection via Pulse Shape Discrimination (PSD). While proving its feasibility, we quantify the magnitude of the gamma-ray rejection in Helium-3 tubes. The implementation of the PSD technique on Helium-3 detectors was also investigated in order to set a possible starting point for the PSD implementation on Boron-10 detectors.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑