专用乳房CT:自适应滤过对剂量分布的影响

Dedicated breast CT: effect of adaptive filtration on dose distribution

Polad M. Shikhaliev

摘要:

目的:本工作的目的是实验研究乳腺剂量分布与自适应过滤。自适应滤波降低了探测器动态范围并提高了图像质量。将具有预定形状的自适应滤波器放置在X射线束处,使得检测器表面处的X射线强度是平坦的。然而,自适应滤波器改变了乳房的平均剂量以及剂量的体积分布。方法:使用直径为14 cm的圆柱形丙烯酸乳房体模测量剂量。制造丙烯酸自适应滤波器以匹配14 cm直径的体模。使用插入沿体模半径从中心到边缘分布的孔中的离子室测量剂量。测量剂量径向分布,用解析函数拟合,计算剂量体积分布和平均值。在40、60、90和120 kVp管电压和6.6 mGy空气比释动能下进行测量。结果:自适应滤波器根据管电压将平均乳腺剂量降低1.6-2.2倍。为了保持图像质量,当使用自适应滤波器时,这些因素增加了管输出。自适应滤波器在40 kVp和60 kVp管电压下提供了均匀的剂量体积分布,但在90 kVp和120 kVp管电压下分布不均匀。结论:虽然自适应滤波有助于探测器动态范围、CT数量均匀性和CT噪声均匀性,但自适应滤波器的剂量体积不均匀性取决于管电压。另一方面,在相同的平均乳腺剂量下,自适应滤波的剂量的体积不均匀性并不高于无自适应滤波。结果还允许确定必要的剂量增加因子,以补偿自适应滤波器中的X射线吸收。

英文摘要:

Purpose: The purpose of the work was experimental investigations of the breast dose distributions with adaptive filtration. Adaptive filtration reduces detector dynamic range and improves image quality. The adaptive filter with predetermined shape is placed at the x-ray beam such that the x-ray intensity at the detector surface is flat. However, adaptive filter alters the mean dose to the breast, as well as volume distribution of the dose. Methods: The dose was measured using a 14 cm diameter cylindrical acrylic breast phantom. An acrylic adaptive filter was fabricated to match the 14 cm diameter of the phantom. The dose was measured using ion chamber inserted into holes distributed along the radius of the phantom from the center to the edge. The radial distribution of dose was measured and fitted by an analytical function and the volume distribution and mean value of dose was calculated. The measurements were performed at 40, 60, 90, and 120 kVp tube voltages and 6.6 mGy air kerma. Results: The adaptive filter decreased mean breast dose by a factor of 1.6-2.2 depending on the tube voltage. To maintain image quality, the tube output was increased by these factors when adaptive filter was used. The adaptive filter provided uniform volume distribution of dose at 40 kVp and 60 kVp tube voltages, but the distribution was non-uniform at 90 kVp and 120 kVp tube voltages. Conclusion: While adaptive filtration helps with detector dynamic range, CT number uniformity, and CT noise uniformity, the volume non-uniformity of the dose with adaptive filter depends on tube voltage. On the other hand, at the same average breast dose, volume non-uniformity of the dose with adaptive filter was not higher that that without adaptive filter. The results allowed also to determine necessary dose increase factor to compensate for x-ray absorption in the adaptive filter.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑