R2 * 弛豫法在复杂多峰多回波化学位移水脂定量中的应用
On the R2* Relaxometry in Complex Multi-Peak Multi-Echo Chemical Shift-Based Water-Fat Quantification: Applications to the Neuromuscular Diseases
摘要:
目的:通过使用针对混杂因素校正并针对信噪比优化的最新理论模型,研究R2* 映射技术的可行性。理论与方法:由于不可忽略的偏差和仍未解决的数值不稳定性,最先进的MRI弛豫算法的性能的改进具有挑战性。在这里,R2* 映射重建,包括复杂的拟合与多光谱脂肪校正,通过使用单衰减(1D)和双衰减(2D)配方,进行了研究,以确定最佳的配置参数,并尽量减少数值伪影。通过模拟和体内数据评价回波数、回波间距和脂肪/水松弛模型的影响。我们还通过分析生物物种之间高百分比的脂肪浸润和局部横向松弛差异的影响来探索这种模型的稳定性。结果:磁共振弛豫测量的主要限制是存在的偏见和伪影的发生,影响其准确性。化学位移复杂的重建R2*-校正与1D配方表现出很大的偏差存在的脂肪和水的弛豫速率的显着差异,脂肪浓度大于30%。我们发现,对于以脂肪为主的组织或受铁过载影响的患者,考虑多指数弛豫时间的MRI重建提供了准确的R2* 测量值,并且不太容易出现数值伪影。结论:在各种诊断情况下,复杂拟合2D公式优于传统的1D近似。虽然它仍然缺乏数值稳定性,需要模型增强和光谱学的支持,它提供了一个可靠的工具,以改善组织表征和监测神经肌肉疾病的松弛测量的发展前景。
英文摘要:
Purpose: Investigation of the feasibility of the R2* mapping techniques by using latest theoretical models corrected for confounding factors and optimized for signal to noise ratio. Theory and Methods: The improvement of the performance of state of the art MRI relaxometry algorithms is challenging because of a non-negligible bias and still unresolved numerical instabilities. Here, R2* mapping reconstructions, including complex-fitting with multi-spectral fat-correction by using single-decay (1D) and double-decay (2D) formulation, are studied in order to identify optimal configuration parameters and minimize numerical artifacts. The effects of echo number, echo spacing, and fat/water relaxation model are evaluated through simulated and in-vivo data. We also explore the stability of such models by analyzing the impact of high percentage of fat infiltrations and local transverse relaxation differences among biological species. Results: The main limits of the MRI relaxometry are the presence of bias and the occurrence of artifacts which affect its accuracy. Chemical-shift complex reconstructions R2*-corrected with 1D formulation exhibit a large bias in presence of a significant difference in the relaxation rates of fat and water and with fat concentration larger than 30%. We find that for fat-dominated tissues or in patients affected by iron overload, MRI reconstructions accounting for multi-exponential relaxation time provide accurate R2* measurements and are less prone to numerical artifacts. Conclusions: Complex fitting 2D formulation outperforms the conventional 1D approximation in various diagnostic scenarios. Although it still lacks of numerical stability which requires model enhancement and support from spectroscopy, it offers promising perspectives for the development of relaxometry as a reliable tool to improve tissue characterization and monitoring of neuromuscular disorders.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。