用置换DOSY(p-DOSY)精确度量失平衡系统的DOSY
Accurate DOSY measure of out-of-equilibrium systems by permutated DOSY (p-DOSY)
摘要:
NMR光谱是监测原位化学反应的极好工具。特别是,DOSY测量非常适合通过确定其尺寸来表征瞬态物质。然而,在这里,我们带来了光的DOSY实验中进行的非平衡系统的困难。在这样的系统中,物质浓度的演变干扰测量过程,并在扩散系数确定上产生偏差,这可能导致错误的解释。我们表明,在DOSY实验过程中使用的一系列梯度强度的随机排列可以平均这种偏差。这种方法,我们将其命名为p-DOSY,不需要改变脉冲序列或处理软件,并完全恢复测量的完全准确性。该技术在监测α-葡萄糖到β-葡萄糖的端基异构化反应上得到了证实。
英文摘要:
NMR spectroscopy is an excellent tool for monitoring in-situ chemical reactions. In particular, DOSY measurement is well suited to characterize transient species by the determination of their sizes. However, here we bring to light a difficulty in the DOSY experiments performed in out-of-equilibrium systems. On such a system, the evolution of the concentration of species interferes with the measurement process, and creates a bias on the diffusion coefficient determination that may lead to erroneous interpretations. We show that a random permutation of the series of gradient strengths used during the DOSY experiment allows to average out this bias. This approach, that we name p-DOSY does not require changes in the the pulse sequences nor in the processing software, and restores completely the full accuracy of the measure. This technique is demonstrated on the monitoring of the anomerization reaction of α- to β-glucose.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。