In vivo evaluation of wearable head impact sensors

Lyndia C. Wu, Vaibhav Nangia, Kevin Bui, Bradley Hammoor, Mehmet Kurt, Fidel Hernandez, Calvin Kuo, David B. Camarillo

摘要:

惯性传感器通常用于测量人类头部运动。一些传感器已经通过假人或尸体实验进行了验证,但缺乏体内评估传感器的方法。在这里,我们提出了一种在体内的方法,使用高速视频来评估牙齿安装(护齿),软组织安装(皮肤贴片),和头盔安装(颅骨帽)传感器在6 - 13克矢状足球头的影响。通过从耳道参考的位移来量化耦合到颅骨的传感器。护齿器位移在视频测量误差范围内(<1 mm),而皮肤贴片和颅骨帽分别从耳道参考位移高达4 mm和13 mm。我们使用距离颅骨位移最小的护齿套作为参考,以评估6自由度皮肤贴片和颅骨帽测量。线性和旋转加速度幅度被皮肤贴片(a_mag的NRMS误差为120%,alpha_mag的NRMS误差为290%)和颅骨帽(a_mag的NRMS误差为320%,alpha_mag的NRMS误差为500%)过度预测。这种过度预测主要是由于平面外运动。为了对传感器误差进行建模,我们发现在前后方向上来自皮肤贴片的面内加速度峰值可以由欠阻尼粘弹性系统来建模。总之,护齿套在体内显示出比其他传感器更紧密的颅骨耦合。此外,在体内的方法是有价值的研究颅骨加速度传感器技术。

英文摘要:

Inertial sensors are commonly used to measure human head motion. Some sensors have been validated with dummy or cadaver experiments, but methods to evaluate sensors in vivo are lacking. Here we present an in vivo method using high speed video to evaluate teeth-mounted (mouthguard), soft tissue-mounted (skin patch), and headgear-mounted (skull cap) sensors during 6-13g sagittal soccer head impacts. Sensor coupling to the skull is quantified by displacement from an ear-canal reference. Mouthguard displacements were within video measurement error (<1mm), while the skin patch and skull cap displaced up to 4mm and 13mm from the ear-canal reference, respectively. We used the mouthguard, which had the least displacement from skull, as the reference to assess 6-degree-of-freedom skin patch and skull cap measurements. Linear and rotational acceleration magnitudes were over-predicted by both the skin patch (with 120% NRMS error for a_mag, 290% for alpha_mag) and the skull cap (320% NRMS error for a_mag, 500% for alpha_mag). Such over-predictions were largely due to out-of-plane motion. To model sensor error, we found that in-plane acceleration peaks from the skin patch in the anterior-posterior direction could be modeled by an underdamped viscoelastic system. In summary, the mouthguard showed tighter skull coupling in vivo than the other sensors. Furthermore, the in vivo methods presented are valuable for investigating skull acceleration sensor technologies.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑