两种不同掩模制备方法对单组分SRS建立不稳定性的评价

Evaluation of Setup Uncertainties for Single-Fraction SRS by Comparing the Two Different Mask-Creation Methods

Jong Geun Baek, Hyun Soo Jang, Young Kee Oh, Hyun Jeong Lee, Eng Chan Kim

摘要:

本研究的目的是基于使用预处理CBCT成像引导的两种不同掩模创建方法的临床数据,评价单次立体定向放射外科(SF-SRS)的设置不确定性。如BrainLAB用户手册所述,为23名患者创建专用无框固定BrainLAB面罩,作为常规面罩(R面罩)制作方法。通过修改患者头部R-面罩的覆盖范围创建的替代面罩(A-面罩)用于23例患者。分析了包括每个掩膜和立体定向目标定位器的系统误差,并将误差计算为LR、SI、AP和偏航设置校正的平均值和标准差(SD)。此外,还分析了三维(3D)向量长度的频率。R掩模在所有方向上的平均设置校正值都很小; < 0.7 mm和< 0.1度,而SD的大小与平均值相比相对较大。与R掩模相比,A掩模的平均值和SD小于R掩模的平均值和SD,AP方向的SD除外。R-面罩和A-面罩系统中偏航旋转方向的平均值和SD相当。对于R掩模,比A掩模更频繁地发生更大幅度的3D矢量移位。立体定向定位系统的每个掩模的设置不确定性具有朝向正AP方向的不对称偏移。A-mask创建方法能够覆盖患者头部的顶部,优于R-mask,因此鼓励SF-SRS使用A-mask以减少设置不确定性。此外,需要仔细的掩模制作以防止可能的设置不确定性。

英文摘要:

The purpose of this study was to evaluate the setup uncertainties for single-fraction stereotactic radiosurgery (SF-SRS) based on the clinical data with the two different mask-creation methods using pretreatment CBCT imaging guidance. Dedicated frameless fixation BrainLAB masks for 23 patients were created as a routine mask (R-mask) making method, as explained in the BrainLAB user manual. The alternative masks (A-mask) which were created by modifying the cover range of the R-mask for the patient head were used for 23 patients. The systematic errors including the each mask and stereotactic target localizer were analyzed and the errors were calculated as the mean and standard deviation (SD) from the LR, SI, AP, and yaw setup corrections. In addition, the frequency of three-dimensional (3D) vector length were also analyzed. The values of the mean setup corrections for the R-mask in all directions were small; < 0.7 mm and < 0.1 degree, whereas the magnitudes of the SDs were relatively large compared to the mean values. In contrast to the R-mask, the means and SDs of the A-mask were smaller than those for the R-mask with the exception of the SD in the AP direction. The mean and SD in the yaw rotational direction in the R-mask and A-mask system were comparable. The 3D vector shifts of a larger magnitude occurred more frequently for the R-mask than the A-mask. The setup uncertainties for each mask with the stereotactic localizing system had an asymmetric offset towards the positive AP direction. The A-mask-creation method, which is capable of covering the top of the patient head is superior to that for R-mask, and thereby the use of the A-mask is encouraged for SF-SRS to reduce the setup uncertainties. Moreover, the careful mask making is required to prevent the possible setup uncertainties.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑