兆伏光子3D打印材料剂量学特性的初步研究
Preliminary Study for Dosimetric Characteristics of 3D-printed Materials with Megavoltage Photons
摘要:
如今,3D打印机在包括放射治疗在内的各个领域都在兴起。这项初步研究的目的是估计3D打印机材料的剂量特性,这些材料可用作辐射治疗中的补偿剂或固定剂。用3D打印机打印出5cm长、密度分别为50%、75%和100%的立方体。使用CT模拟器(Brilliance CT,Philips Medical System,Netherlands)对立方体进行计划CT扫描。计算了6 MV光子束通过立方体后的剂量分布。利用EBT 3薄膜和二维阵列探测器测量了3D打印立方体、空气和水的剂量响应。空气箱结果归一化为100时,TPS计算的剂量与50%和75%立方体的实测剂量为96~99。水和100%立方体的实测和计算剂量为82~84。50%、75%和100%的HU值分别为-910、-860和-10。从这些结果来看,放射治疗中的3D打印机可以准确地用于医疗目的。
英文摘要:
In these days, 3D-printer is on the rise in various fields including radiation therapy. This preliminary study aimed to estimate the dose characteristics of the 3D-printer materials which could be used as the compensator or immobilizer in radiation treatment. The cubes which have 5cm length and different densities as 50%, 75% and 100% were printed by 3D-printer. A planning CT scans for cubes were performed using a CT simulator (Brilliance CT, Philips Medical System, Netherlands). Dose distributions behind the cube were calculated when 6MV photon beam passed through cube. The dose response for 3D-printed cube, air and water were measured by using EBT3 film and 2D array detector. When results of air case were normalized to 100, dose calculated by TPS and measured dose of 50% and 75% cube were 96~99. Measured and calculated doses of water and 100% cube were 82~84. HU values of 50%, 75% and 100% were -910, -860 and -10, respectively. From these results, 3D-printer in radiotherapy could be used for medical purpose accurately.