适用于大量通道的0.042 mm^2可编程双相耳蜗刺激器
A 0.042 mm^2 programmable biphasic stimulator for cochlear implants suitable for a large number of channels
摘要:
本文介绍了一种用于人工耳蜗植入的小型可编程双相刺激器。通过采用双环负反馈,电流发生器的输出阻抗增加,同时最大化输出晶体管的电压顺应性。为了使刺激器电路紧凑,通过使用两级二进制加权晶体管DAC(包括3位高压晶体管DAC和4位低压晶体管DAC)缩放参考电流来设置刺激电流。利用这种结构,可以使电路的功耗和面积最小化。该电路已在AMS 0.18um高压CMOS IC工艺中实现,使用的有源芯片面积约为0.042mm^2。测量结果表明,适当的电荷平衡的阳极和阴极刺激阶段,可以省略隔直电容器。所需面积的减少使得所提出的系统适合于大量的信道。
英文摘要:
This paper presents a compact programmable biphasic stimulator for cochlear implants. By employing double-loop negative feedback, the output impedance of the current generator is increased, while maximizing the voltage compliance of the output transistor. To make the stimulator circuit compact, the stimulation current is set by scaling a reference current using a two stage binary-weighted transistor DAC (comprising a 3 bit high-voltage transistor DAC and a 4 bit low-voltage transistor DAC). With this structure the power consumption and the area of the circuit can be minimized. The proposed circuit has been implemented in AMS 0.18um high-voltage CMOS IC technology, using an active chip area of about 0.042mm^2. Measurement results show that proper charge balance of the anodic and cathodic stimulation phases is achieved and a dc blocking capacitor can be omitted. The resulting reduction in the required area makes the proposed system suitable for a large number of channels.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。