arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 1409.8636physics.ins-det
收藏

多APD氙电致发光器件TPC的研制及特性研究

Development and Characterization of a Multi-APD Xenon Electroluminescence TPC

T. Lux, A. Garcia Soto, O. Ballester, S. Bordoni, I. Gil-Botella, N. Hamer, J. Illa, G. Jover Mañas, C. Martin-Mari, C. Palomares, J. Rico, F. Sanchez, R. Santorelli, A. Verdugo

摘要:

在3.8巴的纯氙压力下,研究了采用多雪崩光电二极管(APD)读出的电致发光(EL)时间投影室(TPC)的性能。相互校准和重建方法的开发和应用的数据产生的能量分辨率为5.3 $\pm$ 0.1半高宽为59.5 keV的伽马从$^{241}$ Am。该结果用基于Geant 4和Penelope的Monte Carlo(MC)再现,其预测所使用的设置的5.2 \% FWHM。APD间距为15 mm时,点分辨率为$\approx 0.5$ mm。这些结果表明,多APD读出是一种有竞争力的技术,为EL探测器填充纯氙。

英文摘要:

The performance of an electroluminescence (EL) Time Projection Chamber (TPC) with a multi avalanche photodiode (APD) readout was studied in pure xenon at 3.8 bar. Intercalibration and reconstruction methods were developed and applied to the data yielding energy resolutions as good as 5.3$\pm$0.1 \% FWHM for 59.5 keV gammas from $^{241}$Am. This result was reproduced with a Monte Carlo (MC) based on Geant4 and Penelope which predicted 5.2 \% FWHM for the used setup. Point resolutions of $\approx 0.5$ mm were obtained with a pitch of 15 mm between the APDs. These results show that multi-APD readout is a competitive technology for EL detectors filled with pure xenon.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑