用于可穿戴心脏监护设备的ECG芯片
An ECG-on-Chip for Wearable Cardiac Monitoring Devices
摘要:
本文介绍了一种用于可穿戴设备中心电信号处理的高集成度、低功耗芯片解决方案。该芯片包含一个具有可编程增益的仪表放大器、一个带通滤波器、一个12位SAR ADC、一个新颖的QRS检波器、8K片内SRAM以及相关的控制电路和CPU接口。模拟前端电路精确地感测原始ECG信号并将其数字化,然后对原始ECG信号进行滤波以提取QRS。使用的采样频率为256 Hz。ECG样本在异步FIFO上进行本地缓冲,并在主机CPU需要时通过SPI接口使用更快的时钟读出。该芯片采用0.35um标准CMOS工艺设计实现。模拟内核工作在1V,而数字电路和SRAM工作在3.3V。芯片总核心面积为5.74 mm^2,功耗为9.6uW。小尺寸和低功耗使该设计适合在可穿戴心脏监测设备中使用。
英文摘要:
This paper describes a highly integrated, low power chip solution for ECG signal processing in wearable devices. The chip contains an instrumentation amplifier with programmable gain, a band-pass filter, a 12-bit SAR ADC, a novel QRS detector, 8K on-chip SRAM, and relevant control circuitry and CPU interfaces. The analog front end circuits accurately senses and digitizes the raw ECG signal, which is then filtered to extract the QRS. The sampling frequency used is 256 Hz. ECG samples are buffered locally on an asynchronous FIFO and is read out using a faster clock, as and when it is required by the host CPU via an SPI interface. The chip was designed and implemented in 0.35um standard CMOS process. The analog core operates at 1V while the digital circuits and SRAM operate at 3.3V. The chip total core area is 5.74 mm^2 and consumes 9.6uW. Small size and low power consumption make this design suitable for usage in wearable heart monitoring devices.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。