从组织近红外衰减数据立方体理论预测适合于空间分辨光谱应用的源-探测器间隔距离
Theoretical prediction of the source-detector separation distance suited to the application of the spatially resolved spectroscopy from the near-infrared attenuation data cube of tissues
摘要:
采用修正的比尔-朗伯定律(MBL)和空间分辨光谱技术,利用连续波近红外光谱技术对肌肉和大脑组织氧化过程进行了测量。空间分辨光谱学通过根据分离测量衰减数据的斜率并基于满足该斜率的线性的分离距离处的衰减斜率计算组织的吸收系数来预测吸收剂浓度的变化。本研究分析了适当的源-探测器的分离距离,通过使用扩散近似分辨率的光子迁移时预测的吸收系数的基础上的组织的反射图像的空间分辨光谱。以吸收系数、约化散射系数和间隔距离为坐标轴,想象三维衰减图像,根据漫反射近似表达式,计算一定坐标区间上的衰减量,得到衰减数据立方体。通过计算和分析衰减相对于坐标的一阶导数和二阶导数,并从衰减数据立方体中计算微分光程因子和衰减相对于吸收系数的一阶导数,我们预测了适合于空间分辨光谱应用的分离距离。分析组织中血红蛋白衍生物时,合适的分离距离为3-5cm,相应的微分光程因子值为3.5 - 5。这些数据与前面的实验数据一致。
英文摘要:
The modified Beer-Lambert law (MBL) and the spatially resolved spectroscopy are used to measure the tissue oxidation in muscles and brains by the continuous wave near-infrared spectroscopy. The spatially resolved spectroscopy predicts the change in the concentration of the absorber by measuring the slope of attenuation data according to the separation and calculating the absorption coefficients of tissue on the basis of the slop in attenuation at the separation distance satisfying the linearity of this slop. This study analyzed the appropriate source-detector separation distance by using the diffuse approximation resolution for photon migration when predicting the absorption coefficient by the spatially resolved spectroscopy on the basis of the reflective image of the tissue. We imagine the 3 dimensional attenuation image with the absorption coefficient, reduced scattering coefficient and separation distance as its axes and obtained the attenuation data cube by calculating the attenuation on a certain interval of coordinate on the basis of the diffuse approximation expression. We predicted the separation distance appropriate for the application of the spatially resolved spectroscopy by calculating and analyzing the first derivatives and second derivatives of attenuation with respect to the coordinates and also doing the differential pathlength factors and first derivatives of the attenuation with respect to the absorption coefficient from the attenuation data cube. When analyzing the hemoglobin derivatives in tissues, the appropriate separation distances are 3-5cm and the value of its corresponding differential pathlength factors are from 3.5 to 5. These data agree with the preceding experimental data.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。