热中子和共振中子对Pu-239 fission室测量快中子注量率的影响
Influence of thermal and resonance neutron on fast neutron flux measurement by Pu-239 fission chamber
摘要:
Pu-239裂变室由于对快中子的平坦响应而被广泛用于测量裂变谱中子通量。然而,同时,共振和热中子可以造成显着的影响,如果他们是慢化,这可以通过使用B和Cd覆盖消除测量。在柱状浓缩铀快中子临界装置上,测量了15 mm厚B盖、0.5mm厚Cd盖和无盖时Pu-239的裂变反应速率分别为1.791 × 10 - 16、2.350 × 10-16和1.385 × 10-15/s。而在20 mm厚的聚乙烯覆盖的裂变室中,Pu-239的裂变反应速率迅速提高到2.569 × 10-14。在1/v和1/E谱假设下,计算了Pu-239热中子和共振中子的平均裂变截面分别为500 b和24.95b,得到热中子、共振中子和快中子通量分别为2.30 × 106、2.24 × 106和1.04 × 108 cm ~(-2)s ~(-1)。
英文摘要:
The Pu-239 fission chambers are widely used to measure fission spectrum neutron flux due to a flat response to fast neutrons. However, in the mean time the resonance and thermal neutrons can cause a significant influence on the measurement if they are moderated, which could be eliminated by using B and Cd covers. At a column enriched uranium fast neutron critical assembly, the fission reaction rates of Pu-239 are measured as 1.791*10-16,2.350*10-16 and 1.385*10-15 per second for 15mm thick B cover, 0.5mm thick Cd cover, and no cover respectively. While the fission reaction rate of Pu-239 is rapidly increased to 2.569*10-14 for a 20mm thick polythene covering fission chamber. The average Pu-239 fission cross-section of thermal and resonance neutrons is calculated to be 500b and 24.95b with the assumption of 1/v and 1/E spectra respectively, then thermal, resonance and fast neutron flux are achieved to be 2.30*106,2.24*106 and 1.04*108cm-2s-1.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。