LiF:Mg,Ti探测器热释光高温比(HTR)与质子能量和剂量的关系
Dependence of the thermoluminescent high-temperature ratio (HTR) of LiF:Mg,Ti detectors on proton energy and dose
摘要:
高温比(HTR)是量化暴露于密集电离辐射后LiF:Mg,Ti辉光曲线高温部分形状变化的参数。引入它是为了估计未知辐射场的有效LET,并校正高线性能量转移(LET)辐射的相对TL效率下降。在目前的工作中,HTR对质子能量(14.5至58 MeV)和剂量(0.5至30 Gy)的依赖关系进行了研究。所有测得的HTR值都在1.2或更高的水平,因此与γ射线的相应值(HTR等于1)显著不同,但发现HTR对20 MeV以上的质子能量的变化不敏感。因此,HTR和相对TL效率之间的关系并不明确。发现HTR依赖于吸收剂量,即使是最低研究剂量。
英文摘要:
The high-temperature ratio (HTR) is a parameter quantifying changes of the shape of the high-temperature part of the LiF:Mg,Ti glow-curve after exposure to densely ionizing radiation. It was introduced in order to estimate the effective LET of an unknown radiation field and to correct the decreased relative TL efficiency for high Linear Energy Transfer (LET) radiation. In the present work the dependence of HTR on proton energy (14.5 to 58 MeV) and dose (0.5 to 30 Gy) was investigated. All measured HTR values were at the level of 1.2 or higher, therefore significantly different from the respective value for gamma rays (HTR is equal to 1), but HTR was found to be insensitive to changes of proton energy above 20 MeV. As a result the relationship between HTR and relative TL efficiency is not unequivocal. The HTR was found to be dependent on absorbed dose even for the lowest studied doses.