摘要:
最近在直线加速器相干光源(LCLS)上实现的双晶金刚石(111)单色器能够将初级X射线光束分成粉红色(透射)和单色(反射)分支。厚度为100 μ m的第一单色晶体提供足够的X射线透射率,以使两条光束线能够同时操作。在这里,我们报告的设计,制造和X射线表征的第一和第二(300微米厚)的晶体中使用的单色器和光学组件持有这些晶体。每个晶体板具有约5 × 2 mm2的区域,具有低缺陷浓度,足以用于LCLS处的X射线光学器件。保持晶体的光学组件被设计成提供在刚性基板上的安装并使安装引起的晶体应变最小化。使用双晶X射线形貌术评估了诱导应变,发现在晶体的5 X 2 mm 2工作区域上诱导应变很小。
英文摘要:
A double-crystal diamond (111) monochromator recently implemented at the Linac Coherent Light Source (LCLS) enables splitting of the primary X-ray beam into a pink (transmitted) and a monochromatic (reflected) branch. The first monochromator crystal with a thickness of 100 um provides sufficient X-ray transmittance to enable simultaneous operation of two beamlines. Here we report on the design, fabrication, and X-ray characterization of the first and second (300-um-thick) crystals utilized in the monochromator and the optical assemblies holding these crystals. Each crystal plate has a region of about 5 X 2 mm2 with low defect concentration, sufficient for use in X-ray optics at the LCLS. The optical assemblies holding the crystals were designed to provide mounting on a rigid substrate and to minimize mounting-induced crystal strain. The induced strain was evaluated using double-crystal X-ray topography and was found to be small over the 5 X 2 mm2 working regions of the crystals.