Multi-sensor microelectrodes for extracellular action potential recording have significantly improved the quality of in vivo recorded neuronal signals. These microelectrodes have also been instrumental in the localization of neuronal signal sources. However, existing neuron localization methods have been mostly utilized in vivo, where the true neuron location remains unknown. Therefore, these methods could not be experimentally validated. This article presents experimental validation of a method capable of estimating both the location and intensity of an electrical signal source. A four-sensor microelectrode (tetrode) immersed in a saline solution was used to record stimulus patterns at multiple intensity levels generated by a stimulating electrode. The location of the tetrode was varied with respect to the stimulator. The location and intensity of the stimulator were estimated using the Multiple Signal Classification (MUSIC) algorithm, and the results were quantified by comparison to the true values. The localization results, with an accuracy and precision of ~ 10 microns, and ~ 11 microns respectively, imply that MUSIC can resolve individual neuronal sources. Similarly, source intensity estimations indicate that this approach can track changes in signal amplitude over time. Together, these results suggest that MUSIC can be used to characterize neuronal signal sources in vivo.
System Identification of a Multi-timescale Adaptive Threshold Neuronal Model
Department of Electrical Engineering and Computer Science, University of Central Florida(电气工程和计算机科学系,中佛罗里达大学);NanoScience Technology Center, University of Central Florida(纳米科学技术中心,中佛罗里达大学)
Generalisation of neuronal excitability allows for the identification of an excitability change parameter that links to an experimentally measurable value