arXiv 1305.6510physics.atom-phastro-ph.IM
收藏

测量氙暗物质探测器中氪污染的原子阱痕量分析系统

An atom trap trace analysis system for measuring krypton contamination in xenon dark matter detectors

Elena Aprile, Taehyun Yoon, Andre Loose, Luke W. Goetzke, Tanya Zelevinsky

摘要:

我们已经开发了一种原子阱痕量分析(ATTA)系统,以测量千分之一万亿分之一(ppt)的水平,与液体氙暗物质探测器超越当前一代的灵敏度实现的先决条件Kr。由于Ar和Kr具有相似的激光冷却波长,因此在测量Kr样品之前已经用Ar测试了该装置以避免污染。射频(RF)等离子体放电产生亚稳态原子束,其使用标准磁光技术进行光学准直、减慢和捕获。根据$1.2 \times 10^{-8}$(检测模式)的测量总系统效率,我们预计ATTA系统在$<2$小时内达到对样品中Kr的ppt浓度的设计目标灵敏度。

英文摘要:

We have developed an atom trap trace analysis (ATTA) system to measure Kr in Xe at the part per trillion (ppt) level, a prerequisite for the sensitivity achievable with liquid xenon dark matter detectors beyond the current generation. Since Ar and Kr have similar laser cooling wavelengths, the apparatus has been tested with Ar to avoid contamination prior to measuring Xe samples. A radio-frequency (RF) plasma discharge generates a beam of metastable atoms which is optically collimated, slowed, and trapped using standard magneto-optical techniques. Based on the measured overall system efficiency of $1.2 \times 10^{-8}$ (detection mode) we expect the ATTA system to reach the design goal sensitivity to ppt concentrations of Kr in Xe in $<2$ hours.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑