A combined ultrasonic flow meter and binary vapour mixture analyzer for the ATLAS silicon tracker
摘要:
ATLAS硅跟踪器冷却控制系统的升级可能需要将C3 F8(八氟丙烷)蒸发冷却剂更改为含有10-25% C2 F6(六氟乙烷)的混合物。这样的变化将降低蒸发温度,以确保在LHC全光度下积累辐射损伤后的热稳定性。此次升级的核心是一种新的超声波仪器,在已知温度和压力下,在流动气体中沿相反方向连续测量声音传播时间,以推断C3 F8/C2 F6流量和混合物成分。本文介绍了该仪器及其监控与数据采集(SCADA)软件。该工具的几种几何形状正在使用或正在评估中。一种用于分析和测量中等流速的具有收缩轴向几何形状的仪器已经证明,对于具有20%C2F6的C3 F8/C2 F6摩尔混合物,混合物分辨率为3.10-3,对于高达30 gs-1的质量流量,流量分辨率为满量程的2%。在分子量(mw)差异很大的混合物中,可以实现更高的混合精度:对于C3 F8泄漏到ATLAS跟踪器氮气外壳的一部分(mw差异160),灵敏度<5.10-5。已开发出一种具有倾斜声路几何形状的仪器,用于约1.2 kg-1的高氟碳质量流率-对应于ATLAS硅跟踪器正在建造的新60 kW热虹吸再循环器中的全流量。进行了广泛的计算流体动力学研究,以确定优选的几何形状(超声换能器间距和位置,以及相对于蒸汽流动方向的声音交叉角)。具有45度交叉角的原型已经证明了对于高达15 ms-1的线性流速,流量分辨率为全尺寸的1.9%。该仪器有许多潜在的应用。
英文摘要:
An upgrade to the ATLAS silicon tracker cooling control system may require a change from C3F8 (octafluoro-propane) evaporative coolant to a blend containing 10-25% of C2F6 (hexafluoro-ethane). Such a change will reduce the evaporation temperature to assure thermal stability following radiation damage accumulated at full LHC luminosity. Central to this upgrade is a new ultrasonic instrument in which sound transit times are continuously measured in opposite directions in flowing gas at known temperature and pressure to deduce the C3F8/C2F6 flow rate and mixture composition. The instrument and its Supervisory, Control and Data Acquisition (SCADA) software are described in this paper. Several geometries for the instrument are in use or under evaluation. An instrument with a pinched axial geometry intended for analysis and measurement of moderate flow rates has demonstrated a mixture resolution of 3.10-3 for C3F8/C2F6 molar mixtures with 20%C2F6, and a flow resolution of 2% of full scale for mass flows up to 30gs-1. In mixtures of widely-differing molecular weight (mw), higher mixture precision is possible: a sensitivity of <5.10-5 to leaks of C3F8 into part of the ATLAS tracker nitrogen envelope (mw difference 160) has been seen. An instrument with an angled sound path geometry has been developed for use at high fluorocarbon mass flow rates of around 1.2 kgs-1 - corresponding to full flow in a new 60kW thermosiphon recirculator under construction for the ATLAS silicon tracker. Extensive computational fluid dynamics studies were performed to determine the preferred geometry (ultrasonic transducer spacing and placement, together with the sound crossing angle with respect to the vapour flow direction). A prototype with 45deg crossing angle has demonstrated a flow resolution of 1.9% of full scale for linear flow velocities up to 15 ms-1. The instrument has many potential applications.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。