弹簧-阻尼系统在临床肺肿瘤运动数据中的应用
Application of a spring-dashpot system to clinical lung tumor motion data
摘要:
建立了一个基于Voigt模型的弹簧-阻尼器系统,用于模拟肺放疗过程中腹部呼吸运动与肿瘤运动之间的相关性。该模型被应用于临床数据,包括52治疗光束从10名患者,三菱实时放射治疗系统,日本札幌。在第1阶段,针对个体患者和射束优化模型参数,以确定参考值并研究模型描述数据的能力。在阶段2中,对于每个患者,将针对单个射束确定的最佳参数应用于来自其他射束的数据,以研究射束特定的模型参数集是否足以对治疗过程中的肿瘤运动进行建模。在第1阶段中,所有单独优化的射束数据的基线均方根(RMS)残差为0.90 ± 0.40 mm。在第2阶段中,发现基于单个射束的患者特定模型参数可以密切模拟肿瘤位置,即使对于不规则射束数据,相对于第1阶段值,RMS误差平均增加0.37 mm。平均而言,获得的肿瘤位置模型输出为95%在阶段1和阶段2中,分别在2.0 mm和2.6 mm的绝对界限内的时间。该模型能够处理输入数据的基线、幅度和频率变化,以及输入肿瘤和输出腹部信号之间的相移。这些结果表明,在首次治疗期间或之前收集患者特异性模型参数,然后在治疗期的剩余时间内保留这些参数可能是可行的。该模型在肺肿瘤放射治疗中具有潜在的临床应用价值。
英文摘要:
A spring-dashpot system based on the Voigt model was developed to model the correlation between abdominal respiratory motion and tumor motion during lung radiotherapy. The model was applied to clinical data comprising 52 treatment beams from 10 patients, treated on the Mitsubishi Real-Time Radiation Therapy system, Sapporo, Japan. In Stage 1, model parameters were optimized for individual patients and beams to determine reference values and to investigate how well the model can describe the data. In Stage 2, for each patient the optimal parameters determined for a single beam were applied to data from other beams to investigate whether a beam-specific set of model parameters is sufficient to model tumor motion over a course of treatment. In Stage 1 the baseline root mean square (RMS) residual error for all individually-optimized beam data was 0.90 plus or minus 0.40 mm. In Stage 2, patient-specific model parameters based on a single beam were found to model the tumor position closely, even for irregular beam data, with a mean increase with respect to Stage 1 values in RMS error of 0.37 mm. On average the obtained model output for the tumor position was 95% of the time within an absolute bound of 2.0 mm and 2.6 mm in Stage 1 and 2, respectively. The model was capable of dealing with baseline, amplitude and frequency variations of the input data, as well as phase shifts between the input tumor and output abdominal signals. These results indicate that it may be feasible to collect patient-specific model parameters during or prior to the first treatment, and then retain these for the rest of the treatment period. The model has potential for clinical application during radiotherapy treatment of lung tumors.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。