用于大电流分流器的宽带数字相位比较器
Wideband digital phase comparator for high current shunts
摘要:
INRIM研制了一种用于大电流分流器相位差精确测量的宽带相位比较器。双输入数字相位检测器通过一对对称的有源保护Transformers连接到比较下的分流器的输出,用精密宽带数字化器实现。数据首先异步获取,然后从板载存储器传输到主机存储器。由于收集了大量数据,因此在采集例程之外执行过滤过程和分析算法。大多数系统误差可以通过适当的反演过程来补偿。该系统适用于在宽电流范围(从几百毫安到100 A)和500 Hz到100 kHz频率范围内比较分流器。用采样频率高达1MHz的数字化仪测量欧洲国家计量机构提供的一组10 A分流器的相位差,在100 kHz以下的频率范围内,扩展不确定度(k=2)小于0.05 mrad。相位比较器的增强版本采用了具有更高采样频率和垂直分辨率的新的数字相位检测器。这允许从20 kHz到100 kHz将对相位检测器的不确定性预算的贡献减少到因子2。理论和实验表明,作为鉴相器耦合的两台高精度宽带数字化仪之间的相位差取决于多个因素,这些因素来自于每个采样系统的模拟和数字印记。
英文摘要:
A wideband phase comparator for precise measurements of phase difference of high current shunts has been developed at INRIM. The two-input digital phase detector is realized with a precision wideband digitizer connected through a pair of symmetric active guarded transformers to the outputs of the shunts under comparison. Data are first acquired asynchronously, and then transferred from on-board memory to host memory. Because of the large amount of data collected the filtering process and the analysis algorithms are performed outside the acquisition routine. Most of the systematic errors can be compensated by a proper inversion procedure. The system is suitable for comparing shunts in a wide range of currents, from several hundred of milliampere up to 100 A, and frequencies ranging between 500 Hz and 100 kHz. Expanded uncertainty (k=2) less than 0.05 mrad, for frequency up to 100 kHz, is obtained in the measurement of the phase difference of a group of 10 A shunts, provided by some European NMIs, using a digitizer with sampling frequency up to 1 MHz. An enhanced version of the phase comparator employs a new digital phase detector with higher sampling frequency and vertical resolution. This permits to decrease the contribution to the uncertainty budget of the phase detector of a factor two from 20 kHz to 100 kHz. Theories and experiments show that the phase difference between two high precision wideband digitizers, coupled as phase detector, depends on multiple factors derived from both analog and digital imprint of each sampling system.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。