利用强度自相关技术模拟研究AGIPD设计选择对X射线光子相关谱的影响

Simulation study of the impact of AGIPD design choices on X-ray Photon Correlation Spectroscopy utilizing the intensity autocorrelation technique

Julian Becker, Christian Gutt, Heinz Graafsma

摘要:

欧洲XFEL,目前正在建设中,将产生一个相干X射线脉冲每222纳秒的脉冲序列高达2700个脉冲。与目前正在开发的快速二维区域探测器相结合,将有可能在亚微秒时间尺度上用非遍历系统进行X射线光子相关光谱学(XPCS)实验。使用探测器模拟工具HORUS对欧洲XFEL采用强度自相关技术的自适应增益积分像素探测器(AGIPD)进行了案例研究。由于当像素尺寸接近(小)散斑尺寸时获得来自XPCS实验的最佳结果,本研究比较了AGIPD(像素尺寸为(200 $\upmu$ m)$^2$)的探测器的可能的有孔版本和具有(100 $\upmu$ m)$^2$像素大小,并研究强度波动和非相干噪声对所获取数据质量的影响。直观的结论是,孔径是没有好处的,因为数据是'扔掉'被证明是正确的低强度。对于强度大于约1光子/(100 $\upmu$ m)发现$^2$开孔是有益的,因为电荷共享效应被排除在外。结果表明,对于所调查的情况,(100 $\upmu$ m)$^2$像素产生的结果显著优于当平均强度超过每(100 $\upmu$ m)$^2$约0.05个光子时,测量(200 $\upmu$ m)$^2$像素。虽然系统在设计上有很大的不同,但它们的信噪比变化仅为2-3倍,提取的相关常数的相对误差变化更小。然而,不同的系统对强度的依赖性表现出明显不同的特点。

英文摘要:

The European XFEL, currently under construction, will produce a coherent X-ray pulse every 222 ns in pulse trains of up to 2700 pulses. In conjunction with the fast 2D area detectors currently under development, it will be possible to perform X-ray Photon Correlation Spectroscopy (XPCS) experiments on sub-microsecond timescales with non-ergodic systems. A case study for the Adaptive Gain Integrating Pixel Detector (AGIPD) at the European XFEL employing the intensity autocorrelation technique was performed using the detector simulation tool HORUS. As optimum results from XPCS experiments are obtained when the pixel size approximates the (small) speckle size, the presented study compares the AGIPD (pixel size of (200 $\upmu$m)$^2$) to a possible apertured version of the detector and to a hypothetical system with (100 $\upmu$m)$^2$ pixel size and investigates the influence of intensity fluctuations and incoherent noise on the quality of the acquired data. The intuitive conclusion that aperturing is not beneficial as data is 'thrown away' was proven to be correct for low intensities. For intensities larger than approximately 1 photon per (100 $\upmu$m)$^2$ aperturing was found to be beneficial, as charge sharing effects were excluded by it. It was shown that for the investigated case (100 $\upmu$m)$^2$ pixels produced significantly better results than (200 $\upmu$m)$^2$ pixels when the average intensity exceeded approximately 0.05 photons per (100 $\upmu$m)$^2$. Although the systems were quite different in design they varied in the signal to noise ratio only by a factor of 2-3, and even less in the relative error of the extracted correlation constants. However the dependence on intensity showed distinctively different features for the different systems.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑