玻耳兹曼常量是通过比较电阻器在水的三相点处的约翰逊噪声与超导约瑟夫森结波形合成器产生的基于量子的电压参考信号来测量的。k = 1.380651(18)\times 10^-23 J/K的测量值与当前CODATA值和组合不确定度一致。这是我们第一次用这种电子技术测量k,也是第一次噪声测温测量,相对综合不确定度为13/10^6。我们描述了我们的约翰逊噪声温度计的最新改进,使统计不确定性的贡献减少到10^6分之七,以及杂散系统误差和EMI效应的进一步减少。详细讨论了这种测量的不确定度预算。

英文摘要:

The Boltzmann constant was measured by comparing the Johnson noise of a resistor at the triple point of water with a quantum-based voltage reference signal generated with a superconducting Josephson-junction waveform synthesizer. The measured value of k = 1.380651(18) \times 10^-23 J/K is consistent with the current CODATA value and the combined uncertainties. This is our first measurement of k with this electronic technique, and the first noise thermometry measurement to achieve a relative combined uncertainty of 13 parts in 10^6. We describe the most recent improvements to our Johnson Noise Thermometer that enabled the statistical uncertainty contribution to be reduced to seven parts in 10^6, as well as the further reduction of spurious systematic errors and EMI effects. The uncertainty budget for this measurement is discussed in detail.