摘要:

我们目前的超低能量的结果与新的球形正比计数器。能量阈值已被推低至约25 eV,并且清楚地收集和检测到单个电子。为了达到这样的性能低能量校准系统已成功开发:-脉冲紫外灯提取光电子从内表面的检测器-各种放射源允许低能量峰值通过荧光过程。基准结果是由于碳荧光而在270 eV处观察到良好分辨的峰,这是这种大质量检测器的独特性能。它为暗物质和低能中微子的研究打开了一扇新的窗口,并可能通过中微子-核弹性散射探测来自核反应堆或超新星的中微子

英文摘要:

We present ultra low energy results taken with the novel Spherical Proportional Counter. The energy threshold has been pushed down to about 25 eV and single electrons are clearly collected and detected. To reach such performance low energy calibration systems have been successfully developed: - A pulsed UV lamp extracting photoelectrons from the inner surface of the detector - Various radioactive sources allowing low energy peaks through fluorescence processes. The bench mark result is the observation of a well resolved peak at 270 eV due to carbon fluorescence which is unique performance for such large-massive detector. It opens a new window in dark matter and low energy neutrino search and may allow detection of neutrinos from a nuclear reactor or from supernova via neutrino-nucleus elastic scattering

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑