基于机器人的远程超声心动图:TER系统在腹主动脉探查中的临床评价

Robot-based tele-echography: clinical evaluation of the TER system in abdominal aortic exploration

Thomas Martinelli, Jean-Luc Bosson, Luc Bressollette, Franck Pelissier, Eric Boidard, Jocelyne Troccaz, Philippe Cinquin

摘要:

目的:TER系统是一种基于机器人的远程超声检查系统。专家在主站点移动超声波探头的模型,机器人再现真实探头的运动,发送回超声波图像和力反馈。该工具可用于在小型医疗中心或偏远地区进行超声检查。本研究的目的是在实际条件下证明TER系统检测腹主动脉瘤和髂动脉瘤的可行性和可靠性。方法:58例患者被纳入法国布雷斯特和格勒诺布尔的2个中心。将远程检查与参考标准床旁检查的主动脉和髂动脉直径测量、动脉瘤检测和描述、动脉粥样硬化检测、检查持续时间和可接受性进行比较。结果:所有动脉瘤(8例)均通过两种技术检测到壁内血栓形成和延伸至髂动脉。主动脉直径的观察者间相关系数为0.982(P <0.0001)。2名术者评价动脉粥样硬化的一致率为84% +/- 11%(95%置信区间)。结论:我们对58例患者的研究表明,TER系统可能是一种可靠、可接受和有效的基于机器人的系统,用于进行远程腹主动脉超声检查。目前正在继续进行研究,以改进一般腹部使用的设备。

英文摘要:

OBJECTIVE: The TER system is a robot-based tele-echography system allowing remote ultrasound examination. The specialist moves a mock-up of the ultrasound probe at the master site, and the robot reproduces the movements of the real probe, which sends back ultrasound images and force feedback. This tool could be used to perform ultrasound examinations in small health care centers or from isolated sites. The objective of this study was to prove, under real conditions, the feasibility and reliability of the TER system in detecting abdominal aortic and iliac aneurysms. METHODS: Fifty-eight patients were included in 2 centers in Brest and Grenoble, France. The remote examination was compared with the reference standard, the bedside examination, for aorta and iliac artery diameter measurement, detection and description of aneurysms, detection of atheromatosis, the duration of the examination, and acceptability. RESULTS: All aneurysms (8) were detected by both techniques as intramural thrombosis and extension to the iliac arteries. The interobserver correlation coefficient was 0.982 (P < .0001) for aortic diameters. The rate of concordance between 2 operators in evaluating atheromatosis was 84% +/- 11% (95% confidence interval). CONCLUSIONS: Our study on 58 patients suggests that the TER system could be a reliable, acceptable, and effective robot-based system for performing remote abdominal aortic ultrasound examinations. Research is continuing to improve the equipment for general abdominal use.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑